© 2016 Thorlabs GmbH
16
MTD415T
MTD415T Data Sheet Rev. 1.0
Command
Explanation
Response Example
Programming
"Ix"
Sets the I Share to
x
*)
Value range
x
:
0
to
100000
[mA/(K+sec)]
Reading
"I?"
Reads the I Share
[x<LF>]
[mA/(K+sec)]
Command
Explanation
Response Example
Programming
"Dx"
Sets the D Share to
x
*)
Value range
x
:
0
to
100000
[(mA*s)/K]
Reading
"D?"
Reads the D Share
[x<LF>]
[(mA*sec)/K]
*) This set value remains during the actual power-on cycle. Use the "
M
" command to save it to the non-volatile
memory.