Plot Environment Data
Beam Stability
The section
Plot Over Time Parameter
is used to configure these plots:
·
The maximum number N of data points in the plots can be defined (N=1 to 100000).
·
The time interval between two data points can be selected (between 0 = every measure-
ment to 1000 s; default = every measurement).
·
Specify the action performed after the maximum number N of recorded data points is
reached:
o
"Discard First Points" means the 1st data set is discarded and the most recent data
point will be added to the plot. In other words, after reaching the maximum number N of
plotted data points, the plot continues and displays the most recent N data points.
o
"Discard Last Points" means that all data points beyond N will be discarded. In other
words, after reaching the maximum number N of data points, the plot stops.
6.5.6 Translation Stage Control
If a translation stage is detected as is the case when a M2MS measurement extension is con-
nected and powered, the topics "Translation Stage Control" and "M
2
set" appear as the last two
topics of the Beam Settings panel. The description in this section assumes a Thorlabs DDS100
Linear Translation Stage, as it is used in the
.
This panel allows the user to manually control the translation stage. This is useful for a coarse
setup of the M² or
.
Actual Position
displays the present stage position.
Target Position:
Enter a value between 0 and 200 mm, press "Enter" on your keyboard, and
the stage will travel with the set
Velocity
(see below) to the new position.
60
61
79
115
Summary of Contents for M2MS
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 14 Feb 2022 ...
Page 73: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 71 Example ...
Page 92: ...BP209 Series 7 Operation Instructions Page 90 21424 D02 ...
Page 119: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 117 ...
Page 158: ...BP209 Series 11 Appendix Page 156 21424 D02 11 10 Drawing M2MS BP209 ...
Page 162: ...BP209 Series 11 Appendix Page 160 21424 D02 11 14 Certifications and Compliances ...
Page 167: ...www thorlabs com ...