BP209 Series
6 The Graphics User Interface (GUI)
Page 32
21424-D02
6.5.3.3 Miscellaneous
Photodiode Bias
Biasing a photo diode decreases its rise time. Set Bias to
On
for measurement of small beam
diameters and M² measurement.
Note
Biasing is possible only for BP209-VIS and BP209IR1 models.
Aperture Width
The maximum aperture diameter is 9 mm. The resolution can be improved if using only 10% of
the aperture for beam analysis, particularly in case of small beam diameters or when using the
the
.
Active Slit Pair
The BP209 Series comes with two pairs of slits. The identification of the slit pairs is given by
their location on the perimeter of the drum, and only the active slit pair is used for beam ana-
lysis.
With
5 µm
slit pair, beam diameters down to 20 µm can be measured using the standard Slit
Scanning method. If the beam diameter is less than that, the Knife Edge Mode must be used
instead, in combination with the
25 µm
slit pair. This is explained in detail in the
.
Scanning Mode
Select
Slit
Scanning for beam diameters > 20 µm (with 5 µm active slit pair) and
Knife Edge
for beam diameters < 20 µm (with 25 µm active slit pair).
6.5.4 Calculation Parameter
This section in the
is used to set parameters for beam profile calculation.
Shown below is the view of the Calculation Parameter section at the first start of the Beam Soft-
ware or after pressing the "
Reset Application Settings to Defaults
" button
in the toolbar:
74
74
26
21
Summary of Contents for M2MS
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 14 Feb 2022 ...
Page 73: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 71 Example ...
Page 92: ...BP209 Series 7 Operation Instructions Page 90 21424 D02 ...
Page 119: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 117 ...
Page 158: ...BP209 Series 11 Appendix Page 156 21424 D02 11 10 Drawing M2MS BP209 ...
Page 162: ...BP209 Series 11 Appendix Page 160 21424 D02 11 14 Certifications and Compliances ...
Page 167: ...www thorlabs com ...