Rev K Jan 2020
Page 14
Chapter 4 Software & Electrical Installation
4.2 Electrical Installation
4.2.1 Connecting To The Supply
4.3 Connecting The Hardware
1) Perform the mechanical installation as detailed in Section 3.
2) Install the APT Software - see Section 4.1.
3) Using the USB cable provided, connect the stage unit to your PC.
4) Connect the stage to the power supply - see Section 4.2.1.
5) Connect the PSU to the main supply and switch ‘ON’.
6) Wait for the unit to initialize (about 5 sec). Do not press any controls during this
time. The ENABLE LED is lit when the unit is ready for use.
7) Windows
TM
should detect the new hardware. Wait while Windows
TM
installs the
drivers for the new hardware - see the Getting Started guide for more information.
Fig. 4.1 Power and USB Connectors
Warning: Risk of Electrical Shock
The PSU unit must be connected only to an earthed fused supply of 110 to 230V.
Use only the power supply supplied by Thorlabs, other units may not be rated to
the same current.
The unit is shipped to the UK, Europe and the USA, with the appropriate power
plug already fitted. When shipped to other territories the appropriate power plug
must be fitted by the user. Cable identification is as follows:
Brown: Live Blue: Neutral Green/Yellow: Earth/Ground
When connecting the power and USB cables, ensure they are routed clear of the
moving platform.
Note
The USB cable should be no more than 3 metres in length. Communication lengths
in excess of 3 metres can be achieved by using a powered USB hub).
_
+