![THORLABS LDM9T User Manual Download Page 20](http://html.mh-extra.com/html/thorlabs/ldm9t/ldm9t_user-manual_1110765020.webp)
LDM9T
Chapter 8: Specifications
Page 18
15827-D02
RF Modulation
RF modulation can be accomplished in one of two ways. For low frequencies
(below ~200 kHz), the RF modulator should be connected to the LD controller.
Figure 10 shows the performance using the LDC202C. Above 200 kHz, the
modulator should be connected to the mount’s SMA connector. The mount has
been tested up to 1 GHz, (see Figure 11 below).
Figure 10 Low Frequency Modulation Through LD Controller
Figure 11 RF Modulation Through SMA Connector on LM9LP