Specification
Description
Electrical Characteristics
Max Input Voltage Range
-10 to 10 V
Input Impedance
1 k
Output Voltage
-200 to 200 V
Output Impedance
50
Slew Rate
400 V/
s
Output Noise
1.5 mV RMS
Voltage Gain
1
-20 ± 2%
DC Bias Adjust
-200 to 200 V
HV Monitor to Output Ratio:
With Input Impedance of 50Ω
40:1 (Vout / 40 ± 6%)
With Input Impedance of >10 kΩ
20:1 (Vout / 20 ± 6%)
HV Monitor Output Impedance
50
AC Power
115 V/230 V, 50 – 60 Hz, 70 VA
Max Ratings
Max Output Current
100 mA DC
Max Input Voltage Range
-10 to 10 V
Fuse Rating
630 mA @ 115 VAC (5x20 mm SLO-BLO)
400 mA @ 230 VAC (5x20 mm SLO-BLO)
Operating Temperature Range
10 to 40 °C, Max 85% RH
Physical Features
Input Connector
BNC (± 10 V, 10 mA)
HV Output Connector
2
BNC (± 200 V, 100 mA)
HV Monitor Connector
3
BNC (± 10 V, 200 mA, Min Load 50 Ω)
Bias Adjustment
Digital Encoder
Output Enable
Front Panel Pushbutton
Output HV Indicator
Bright LED
Power Switch
Rocker Switch
Dimensions
9″ x 5″ x 12.5″
228.6 mm x 127 mm x 317.5 mm
Weight
11.6 lbs
Other
Tilting Rubber-Padded Feet
1
The voltage gain is inverted to preserve the high slew rate of the output amplifier (i.e., a -1 V input results in
+20 V output).
2
The HV output will handle an accidental short circuit without damage, but the output should not be shorted
continuously.
3
The HV monitor output will handle accidental short circuits without damage, but must not be shorted for more
than 5 sec to prevent excessive heating of the output resistor. The HV monitor output will source up to 200 mA
when a short circuit is applied.