Press to turn on the instrument. Hold the instrument within 1.5 inches from the center of the forehead and press
and hold ‘Start’ until you hear one beep (unless muted) to indicate the reading is complete. Readings should take
approximately 1 second.
The instrument will automatically turn off after 60 seconds. Manually turn the instrument off by pressing and holding the
for 5 seconds.
The temperature displayed is the “oral equivalent” value. If the temperature is above 99.5°F (37.5°C) you will hear three
short beeps followed by one long beep.
For Best Results:
Keep the forehead area clean. Keep away from sweat, cosmetics, and scars while taking measurements.
Please remain in a stable environment for 5 minutes and avoid exercising, bathing, or showering for 30 minutes,
which can artificially raise your temperature.
Fever alarm
Surface temperature
Forehead temperature
Mute mode
Memory locations
Features
Operating Instructions for Forehead Use
Indications for Use
Specifications
ThermoWorks WAND is a non-contact digital forehead thermometer intended for the intermittent measurement of human
body temperature in people of all ages.
Forehead Range
Surface Range
Operating Range
Storage Range
Transport. Temp
Atmospheric Pressure
Forehead Accuracy
Surface Accuracy
Resolution
Units
Response Time
Auto-Off
Battery
Dimensions
Weight
Expected Service Life
Contents
93.2 to 108°F (34 to 42.2°C)
–7.6 to 176°F (–22 to 80°C)
50 to 104°F (10 to 40°C) 15 to 85% RH
–4 to 122 (–20 to 50°C), RH
≤
85%
Shall be less than 158°F (70°C), RH
≤
95%
800 to 1013 hPa
±0.4°F (0.2°C) within 95 to 107.6°F (35 to 42°C) Otherwise ±0.5°F (0.3°C)
±0.5°F (±0.3°C) with 17.6 to 108°F (22 to 42.2°C) Otherwise ±4% or ±4°F (±2°C), whichever is greater
0.1°
°F/°C
1 second
60 seconds
2 x AAA, 3,000 hours
6.2 L x 1.9 W x 1.6 D inches (158 L x 48 W x 40.2 D mm)
3.5 oz (100 g)
4 years
Thermometer x 1, Operating Instructions x 1, Battery AAA x 2
START
Stand-by
(ON/MEM)
LCD
Lens
Battery Cover
THD2FE
28.9 mm
14
.6 m
m
18
.1 m
m
24.1 mm
1639
LCD
Stand-by
(ON/MEM)
START
Probe Lens
Battery Cover