ThermoWorks Thermapen Mk4 Manual Download Page 2

Making the Thermapen Trim Adjustment:

After determining that your Thermapen requires adjustment by following the 
above procedure, follow these steps to trim the Thermapen:

1.  Turn the Thermapen off by folding the probe away and 

locking it into the probe retaining slot.  

2.  Calculate the trim adjustment value by subtracting the reading 

written down in Step 3 of ‘Determining if Your Thermapen Requires 
Adjustment’ from 32.0°F (or 0.0°C).

Example:

•  Reading from my Thermapen = 32.9°F
•  Adjustment Value:  32.0°F - 32.9°F = –0.9°F 

3.  Using a Philips screwdriver, remove the battery compartment to access the Set and Menu buttons. 

4.  Open the probe to turn on the Thermapen. 

5.  Hold down the menu button for 5 seconds until ‘SET TRIM’ is displayed. Press the Set button until the target trim 

adjustment value is displayed. Note the trim adjustment value range is from –3.6°F to +3.6°F (-2°C to +2°C). 

6.  Press the Menu button to save the trim setting, ‘END’ is displayed for 1 second before returning to normal operation.

7.  Retest in the ice bath to confirm that the new reading is within the acceptable limits: between 31.3°F and 32.7°F (–0.4°C 

and +0.4°C). If the reading is still outside of the acceptable limits, redo the trim adjustment by following steps 1 through 7.

For further technical assistance, contact Technical Support @ 801-756-7705 or [email protected].

ThermoWorks, Inc

  •  741 E. Utah Valley Dr., American Fork, UT 84003  • ph: 801-756-7705  •  fax: 801-756-8948  •  [email protected]  •  

www.thermoworks.com

P-15-007-M-2-a

Thermapen

®

 Mk4 

Trim Adjust Procedure

(continued)

SET MENU

SET MENU