10
6.
Open the reservoir and add additional fluid as needed. Close the cap securely to the reservoir.
7.
It is possible that the initial startup and fluid transport could generate a “Low Fluid” alarm.
Clear the alarm by pressing the “Accept” key. Open the reservoir cap and add fluid. Close the
cap and press “Start” to re-‐start the chiller.
8.
When the unit is operating and controlling to set temperature, the RUN Mode screen content
will be displayed on the LCD screen.
Note:
If the application is located above the chiller, there is a potential for the fluid to overflow the
reservoir due to gravity, when the cap is opened with the pump stopped. During these cases, it is
recommended that the external tubing sections be individually primed by connecting it to the chiller.
When all sections are primed, then connect the external tubing to the chiller and the application, and
complete the final priming of the application.