Error Codes 

ErrHI

 indicates the sensor is operating above its temperature limit, the Probe Assembly is open circuited from a cut or broken 

section, or the cable is not properly inserted at the circuit board terminal. The most common cause is a damaged cable. 

 

ErrLO

 indicates the sensor is operating below its temperature limit, the Probe Assembly is short circuited due to a smashed or cut 

section, or the cable wire polarity is reversed at the circuit board terminal. The most common cause is a damaged cable. 

 
nO rEAd

 -the user accessed "List" or "At" before saving temperatures. 

 

 

 

REPLACING THE PROBE ASSEMBLY 

 

NOTES:  

1)

 

Replacement of the probe assembly requires re-calibration of the device. Replacement should only be done by experienced 
personnel and if calibration equipment is available.

  

2)

 

Please refer to IEC/EN 60079-19 when making the repair 

3)

 

Only use replacement probe assemblies obtained from ThermoProbe, Inc. or one of its authorized distributors with part 
specifications as follows:

 

 

 
a) First follow REPLACING BATTERY instructions a through c to remove batteries.  
b) On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement.  
 

White – positive sensor wire 

 

Black – negative sensor wire and sheild 

c) Set the cover and circuit board aside and remove the strain relief knot in cable assembly.  
d) Unwrap the cable from the assembly and pull the cable free of the rubber grommet.  
e) Insert the new cable wire through the rubber grommet and then pull several inches of cable slack past the grommet.  
f) Tie a simple overhand knot in the cable at the grommet for strain relief and pull any slack through the grommet.  
g) On the circuit board, push the terminal clamps down and insert the new wire leads making sure the black wire lead is installed in 
the terminal indicated with black paint. (See Figure 1) 
h) Reinstall the batteries and cover and re-spool the cable assembly.  
i) Perform a calibration.  
 

 

Negative Terminal 
indicated by black 
marking 

Summary of Contents for TP7C

Page 1: ...ry risks e Power source must be removed before performing any maintenance f Exchange of components other than the batteries may compromise ATEX IECEx certification and shall only be undertaken by Thermoprobe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the tank according to clause 6 3 2 e IEC EN...

Page 2: ...rdous area to a level below that which will ignite the gases In order to have a fire or explosion fuel oxygen and a source of ignition must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never be great enough to cause ignition TP7C TP9 Hazardou...

Page 3: ...onic AA LR6 Alkaline LR6XWA GP Gold Peak AA LR6 Alkaline GP15A MEASURING PROCEDURES See www thermoprobe net for video on proper use of this instrument Contact your distributor or ThermoProbe Inc for names of training facilities Refer to American Petroleum Institute measurement standard Chapter 7 2 When the temperature has stabilized both Up and Down arrows will be displayed If you wish to log the ...

Page 4: ...te Lo bAtt on the display before resuming normal display functions When the batteries are low the backlight will not operate in order to conserve power while the user completes his operations Replace batteries as soon as possible in a safe location after Lo bAtt is noticed this will ensure backlight operation and avoid possible malfunctioning Do not attempt to calibrate the instrument if the Lo bA...

Page 5: ...ained from ThermoProbe Inc or one of its authorized distributors with part specifications as follows a First follow REPLACING BATTERY instructions a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement White positive sensor wire Black negative sensor wire and sheild c Set the cover and circuit board asi...

Page 6: ...ndations Calibration must not be performed in any environment considered to be hazardous Equipment needed Ice Bath or other low temperature bath with certified reference thermometer Warm to hot fluid bath between 20 C approx 68 F or higher up to 90 C approx 194 F with certified reference thermometer see Note Optional high temperature oil bath at about 150 C 300 F and certified reference thermomete...

Page 7: ...ust the device to the actual temperature The On button will decrement the readings and the f Button will increment the reading Once the device temperature matches the actual temperature press the Calibration Button to save the setting The display will scroll SAUE or CAnCEL If the SAUE option is chosen the blinking letter will change to represent the next temperature level A b If the CAnCEL option ...

Reviews: