6  UNDO-CALIBRATION

If you want to undo the calibration and 
go back to the default calibration 
setting, please follow the below steps:

1. Turn on the thermometer and press 
    the “CAL” button 5 seconds  to enter 
    Calibration mode;

2. Press the lock button to undo the 
    calibration and it will reset to the 
    default setting.

7  HELPFUL HINTS

When measuring different foods or 
meat, it may be desirable to cool the 
probe first by rinsing the tip with 
cold water.
It is recommended that you wipe the 
probe tip clean between readings. 
This is especially important when 
taking temperature readings from 
different foods or meat.

The thermometer is designed for 
temporary, hand held use. DO NOT 
use in an oven. DO NOT leave it in 
extremely hot foods or liquid

-6-

-7-

EN

EN

5. Turn on the thermometer, unscrew 
    and take off  the battery cover.

6. Immerse the probe in the ice/water 
    mixture and make sure not to touch 
    the sides of the container. Wait till 
    the temperature reading drops to its 
    lowest value, which should be around 
    32°F or 0°C. If the temperature of the 
    ice/water mixture is not within 28°F 
    to 36°F (or -2°C to +2°C), the 
    calibration can’t be continued: you 
    will need to follow the above steps to 
    re-make the ice/water mixture;

7. Press and hold the CAL button for 5 
    seconds to enter the Calibration 
    mode. CAL will display for 3 seconds 
    and current temperature will display 
    on the screen. 

8. Press the CAL button once to confirm 
    the calibration mode and exit the 
    calibration mode.

Note:

The Calibration mode will only last for 
15 seconds. If you’re unable to finish 
the calibration within 15 seconds, the 
thermometer will automatically exit 
Calibration mode.