
8
Maintenance
Preventative maintenance consists of inspection, testing, checking connections, and
general cleaning of equipment at scheduled intervals. The maintenance recommen-
dations that follow are intended to support and in some cases “add to” those procedures
detailed in the facility’s Planned Maintenance System (PMS). In case of conflicts, con-
tact the project engineer for resolution. When carrying out the scheduled maintenance
program, the following safety precautions should be observed.
Safety Precautions
The heat tracing can be powered by the project specified nominal voltages ranging from
100 to 600 Vac. It is important that only authorized trained personnel conduct these
maintenance and service activities. Before conducting any maintenance or service pro-
cedure, exercise required lockout and tag out procedures at the appropriate circuit break-
ers. Additionally, do additional testing within the control panel to ensure that the specific
heat tracing and control circuit of interest is fully de-energized and the equipment is
grounded.
If it becomes necessary to service or test live equipment, the following instructions must
be followed:
• Use one hand when servicing the equipment. Accidental death or severe injury
may occur especially if a current path is created through the body from one hand
to the other.
• First, de-energize the equipment. To de-energize any capacitors connected into
the circuits, temporarily ground the terminals where work is to be done.
• Connect the multi-meter/instrument to the terminals of interest using a range
higher than the expected. Make sure that you are not grounded whenever a
need arises to adjust equipment or test circuit operation. Verify that all test
equipment used is properly maintained and safe for the intended use.
• Without touching the multi-meter/instrument energize the equipment and read
the values indicated on the multi-meter/instrument.
• Remove the test leads after de-energizing the circuit of interest.
24
PN 50312_0414
Summary of Contents for TraceNet TCM18
Page 32: ...9 Notes 29 PN 50312_0414 ...