‘STAR-700’ Operating Guidelines with 25 Segment Programmer.
Safe Operating Value (SOV).
On initial Power up the STAR-700 Touchscreen shows the ‘Main Screen’ and controls the ‘Safe
Operating Values’ only. You will notice that the indicator in the top right corner states that the
Program Cycle is OFF. The Values below the heading ‘Safe Operating Value’ are the current set
values and the controller will start to control these values. To change the ‘Safe Operating Value’
simply touch the ‘Operating Value’ box. A Pop-Up numeric keypad is displayed. Type in the new
value and press enter. The cabinet will automatically control the new ‘Safe Operating Value’.
The ‘Safe Operating Value’ has been designed to be set by the operator to a safe/normal condition
that will not cause damage to any Research, Experimental, Plant Material and/or any items being
stored in the cabinet in the event of a power failure. If a ‘Program Cycle’ is operating and a power
failure occurs, on resumption of power the controller will revert to its ‘Safe Operating Values’.
Main Menu Screen (SOV Mode)