![ThermoFisher Scientific thermo scientific imPULSE S.U.M. User Manual Download Page 123](http://html1.mh-extra.com/html/thermofisher-scientific/thermo-scientific-impulse-s-u-m/thermo-scientific-impulse-s-u-m_user-manual_1105607123.webp)
Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
118
5. Hand-tighten the adapter and verify that the probe tip is not touching
the membrane.
6. Place the probe assembly with probe into the autoclave tray for
probe kits (Figure 5.30).
Figure 5.30. Probe assembly and autoclave tray.
Chapter 5
|
BPC loading instructions
7. Autoclave the probe assembly using a validated sterilization cycle
(approximately 30 minutes at 122°C). A 30-minute sterilization cycle
is generally sufficient. Options of wet or dry cycle parameters can
be used. Slow exhaust cycles are preferred, as this minimizes stress
on the probes during the temperature and pressure changes of
autoclaving. Many dry cycle parameters are not slow exhaust.
8. Allow sufficient time for the probe assembly to cool completely
before connecting to the BPC for probe insertion.
9. When stored properly, autoclaved probe assemblies can be stored
dry for short periods of time (less than 24 hours) without loss of
sensor longevity, performance, or sterility.
5.3.2 Making CPC AseptiQuik connections
CPC AseptiQuik G genderless connector components
Figures 5.31 and 5.32 below illustrate the components of CPC
™
AseptiQuik
™
G genderless connectors. Connectors with white
protective cover pull tabs may be autoclaved. Generally, connectors
with blue pull tabs are gamma irradiated, not autoclaved. Visit the
Colder Products Company website at http://cpcworldwide.com for
more information.
For instructions on making an aseptic connection, see the following
section.