imPULSE S.U.M. with Touchscreen Console User’s Guide
|
111
Chapter 5
|
BPC loading instructions
5. Remove other packing materials (Figures 5.8–5.10) except for the
C-clips (Figure 5.11). These are sections of tubing on the edge of
the mixing disc that provide protection during BPC loading.
Note:
Be sure to remove the round, closed-cell foam from the folds of the
BPC, shown in Figure 5.8.
Figure 5.8. Removing round foam
from BPC folds.
Figure 5.9. Removing foam
packaging.
Figure 5.10. Removing bubble
wrap from tubing sets.
Figure 5.11. C-clips on the edge of
the mixing disc.
5.2.2 Loading the BPC
Standard loading
1. Use both hands (or two operators as needed) to lift and place the
BPC into the unit (Figure 5.12). The shaft should be facing down,
with the two outlet tubes in the front (facing you). Tilt the BPC to
allow access to the mixer shaft (Figure 5.13), and place the BPC
into the vessel (Figure 5.14).
Note:
For 30 L to 100 L units, load the
BPC through the top of the vessel instead of the front vessel door.