Pressing the POWER ON button should not allow power ing ON the Equinox.
•
The emergency stop button has to be unlocked with the key.
8.5 Replacing the Main Power Supply Cord
If the shipped power supply cord has to be replaced, the new power supply cord must comply
with following specifications:
•
250VAC 10A or 125VAC 10A
•
Cross section of wires: 3x0.75 mm²
•
IEC 60320/C13 plug type
8.6 Replacing the Fuses
The fuseholder is located below the appliance inlet. To access the fuses, use a screw driver
to pull the tray with the fuses.
Two fuses are used and with following specifications:
Size 5x20mm
4A fuse
Ratings 250VAC 1500A
Time-Lag T
This type of fuse is manufactured for example by Schurter under the r eference 0001.2510.
8.7 Cleaning
External parts and the enclosure can be cleaned with a soft and dry duster. Domestic alcohol
is forbidden as it can damage paint and markings.