background image

 

Operation

Aligning the Centrifuge

Thermo Scientific

Legend XF / XFR

4-13

Aligning the Centrifuge

1. To turn off the centrifuge put the mains switch to "0". 

Summary of Contents for Thermo Scientific 7500 4532

Page 1: ...Thermo Scientific Sorvall Legend XF XFR Centrifuge Instruction Manual Visit us online to register your warranty thermofisher com labwarranty 50120960 f 03 2020...

Page 2: ...WEEE Conformity This product is subject to the regulations of the EU Waste Electrical Electronic Equipment WEEE Directive 2002 96 It is marked by the following symbol...

Page 3: ...apter 2 Before use 2 1 Before Setting up 2 2 Setting up 2 2 Transporting the Centrifuge 2 4 Location 2 4 Aligning the Centrifuge 2 4 Mains Connection 2 5 Storage 2 6 Chapter 3 Control Panel 3 1 Contro...

Page 4: ...Cleaning Intervalls 6 2 Cleaning 6 2 Disinfection 6 3 Decontamination 6 4 Autoclaving 6 5 Service of Thermo Fisher Scientific 6 5 Chapter 7 Trouble Shooting 7 1 Mechanical Emergency Door Release 7 2 T...

Page 5: ...uge is an IVD accessory and therefore subject to the Directive 98 79 EC of the European Parliament and the Council of October 27 1998 on in vitro diagnostic medical devices This centrifuge is a labora...

Page 6: ...o be used for its intended use only Plug the centrifuge only into sockets which have been properly grounded Do not move the centrifuge while it is running Do not lean on the centrifuge Use only rotors...

Page 7: ...d react violently with one another The centrifuge is neither inert nor protected against explosion Never use the centrifuge in an explosion prone environment Do not centrifuge inflammable substances D...

Page 8: ...l hazards CAUTION means that material damage could occur WARNING means that injuries or material damage or contamination could occur This symbol refers to biological hazards Observe the information co...

Page 9: ...on and Description Contents Characteristics of the Sorvall Legend XF XFR on page 1 2 Technical Data on page 1 3 Directives Standards and Guidelines on page 1 4 Functions and Features on page 1 5 Mains...

Page 10: ...nd RCF or the entry mode These settings can be changed even while the centrifuge is running With the help of the PULSE key you can also centrifuge a sample for just a few seconds if called for The Sor...

Page 11: ...II II Pollution degree 2 2 Heat dissipation ventilated refrigerated 230V 120V 230V 120V 6653 BTU h 4776 BTU h 5800 BTU h 4776 BTU h IP 20 20 running time unlimited unlimited max Speed nmax 15200 rpm...

Page 12: ...2 Overvoltage category II IEC 61010 2 101 Emitted interference FCC Part 15 CLASS A NOTE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Par...

Page 13: ...data is saved Functions RCF selection temperature control pretemp with cooling equipped devices Acceleration braking profiles 9 acceleration and 10 braking curves Rotor recognition Automatic Imbalance...

Page 14: ...ating manuals For more information visit our website at http www thermofisher com TX 750 7500 3607 with round buckets 7500 3608 with rectangular buckets 7500 3614 incl Microtiter carrier 7500 3617 inc...

Page 15: ...Legend XF XFR 2 1 2 Before use Contents Before Setting up on page 2 2 Transporting the Centrifuge on page 2 4 Location on page 2 4 Aligning the Centrifuge on page 2 4 Mains Connection on page 2 5 Sto...

Page 16: ...ny and Thermo Fisher Scientific immediately if any damage is discovered 2 Remove the packaging 3 Check the order for completeness see Scope of Supply on page iii If the order is incomplete please cont...

Page 17: ...the rails onto the pallet in front of the centrifuge 4 Remove the protectors that locate the centrifuge on the pallet Release the castors Use several people in order to roll the centrifuge off the pa...

Page 18: ...ree of resonance for example a level laboratory bench The supporting structure must be suitable for horizontal setup of the centrifuge The centrifuge should not be exposed to heat and strong sunlight...

Page 19: ...ch on the back press 0 2 Plug the centrifuge into grounded electrical sockets only 3 Check whether the cable complies with the safety standards of your country 4 Make sure that the voltage and frequen...

Page 20: ...Storage Before storing the centrifuge and the accesories it must be cleanedand if necessary desinfected and decontaminated Store the centrifuge in a clean dust free location Be sure to place the cent...

Page 21: ...Thermo Scientific Legend XF XFR 3 1 3 Control Panel Contents Control Panel on page 3 2 Keys on page 3 2...

Page 22: ...tents Start Normal start of the centrifuge Stop End run manually Open lid Automatic release possible only when device is switched on Emergency release see Mechanical Emergency Door Release on page 7 2...

Page 23: ...ge 4 2 Lid Opening on page 4 2 Rotor Installation on page 4 2 Entering Parameters on page 4 4 Centrifugation on page 4 10 Temperature Adaptation during Standstill on page 4 11 Short term Centrifugatio...

Page 24: ...hows The top line shows the message Lid open The bottom line shows the values for the last run Lid Opening 1 Press the OPEN key Close Lid 2 Close the lid by pressing down on it lightly in the middle o...

Page 25: ...s properly installed by lifting it slightly on the handle If the rotor can be pulled up then it must be reclamped to the centrifuge spindle CAUTION Unapproved or incorrectly combined accessories can c...

Page 26: ...ed profile by entering a digit between 1 9 on the numeric pad Curve number 1 is the slowest and curve number 9 the fastest 3 Confirm your entry by pressing ENTER Braking curve 1 Press the ACC DEC key...

Page 27: ...corrected automatically if the resulting speed is less than 300 rpm This is because 300 rpm is the lowest selectable speed Explanation of RCF Value The relative centrifugal force is given as a multip...

Page 28: ...nning until you stop it manually with the STOP key Run time mode Use the TOGGLE key to set the run time counter start AT START When AT START is lit the running time will be counted right from pressing...

Page 29: ...desired temperature using the numeric pad The digits show in sequential order 3 Confirm your entry by pressing ENTER Your entry will be automatically confirmed if you do not press any key for 5 second...

Page 30: ...igits of the bucket catalog number 1 Press the BUCKET key The display shows the following message 2 Press the BUCKET key repeatedly until the bucket being used is displayed Your entry will be automati...

Page 31: ...a program Programs can be stored and retrieved later by means of the assigned number 1 Enter the program parameters as described above 2 Press the program memory key to load any of the programs 1 5 Pr...

Page 32: ...ours ago this message will apear in the display Check the castor breaks before pressing the START key again Starting centrifuge program Press the START key on the control panel The centrifuge accelera...

Page 33: ...ed at the designated rate The message END will illuminate and after pressing the key OPEN the lid will open and you can remove the centrifuged material Temperature Adaptation during Standstill The tem...

Page 34: ...ge spindle Make sure not to tilt the rotor while doing this Aerosol tight Rotors When using an aerosol tight lid the rotor can only be removed with the lid closed This is to protect you and the sample...

Page 35: ...4 Operation Aligning the Centrifuge Thermo Scientific Legend XF XFR 4 13 Aligning the Centrifuge 1 To turn off the centrifuge put the mains switch to 0...

Page 36: ...Thermo Scientific Legend XF XFR 5 1 5 System Menu...

Page 37: ...7 Deutsch 1 English 2 Deutsch 3 Fran aise 4 Espa ol 5 Italiano 6 Nederlands 7 STOP ENTER LCD powersave START 0 no 1 yes ENTER ENTER No Program passcodes 0 0547V011 1 Software and NVRAM identification...

Page 38: ...n Otherwise press 0 no Press ENTER key to confirm Keypad beep Press 1 yes on the numeric pad if the centrifuge should make beep when pressing any key Otherwise press 0 no Press ENTER key to confirm LC...

Page 39: ...1 yes on the numeric pad if the centrifuge should open after the run Otherwise press 0 no Press ENTER key to confirm Software ID Here you fine the current software version Press ENTER key to confirm C...

Page 40: ...d XF XFR 6 1 6 Maintenance and Care Contents Cleaning Intervalls on page 6 2 Cleaning on page 6 2 Disinfection on page 6 3 Decontamination on page 6 4 Autoclaving on page 6 5 Service of Thermo Fisher...

Page 41: ...ver exceed 50 C since higher temperatures could damage the material and shorten the lifetime of the parts Use only disinfectants with a pH of 6 8 Dry aluminum parts off with a soft cloth After cleanin...

Page 42: ...centrifugation The rotor chamber and the rotor should be treated preferably with a neutral disinfectant CAUTION Before using any cleaning or decontamination methods except those recommended by the ma...

Page 43: ...inet at a maximum temperature of 50 C After cleaning treat the entire surface of aluminum parts with corrosion protection oil 7000 9824 Also treat the cavities with oil Tread the bold of the swing out...

Page 44: ...parts with corrosion protection oil 7000 9824 Also treat the cavities with oil Tread the bold of the swing out rotor with bold grease 75003786 Autoclaving 1 Before autoclaving clean rotor and accesso...

Page 45: ...ic Thermo Fisher Scientific offers inspection and service contracts for this work Any necessary repairs are performed for free during the warranty period and afterwards for a charge This is only valid...

Page 46: ...Thermo Scientific Legend XF XFR 7 1 7 Trouble Shooting Contents Mechanical Emergency Door Release on page 7 2 Troubleshooting by User on page 7 3 When to contact Customer Service on page 7 5...

Page 47: ...g process lasts much longer than usual Proceed as follows 1 Make sure the rotor has stopped view port in the lid 2 Pull out the power supply plug 3 On the back of the housing are two black plastic plu...

Page 48: ...ed Bucket code is undefined for the rotor detected check the set points for the given bucket code Is it permitted to use the current bucket in the rotor currently mounted Restart the centrifuge If an...

Page 49: ...e If an error message appears again inform Customer Service E 13 The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked The check sum in the data mem...

Page 50: ...trifuge If an error message appears again inform Customer Service E 33 The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked Overpressure in the ref...

Page 51: ...en to contact Customer Service 7 6 Legend XF XFR Thermo Scientific 3 Press and hold the ENTER key until the following message is displayed 4 Communicate the software version to the service technician...

Page 52: ...lcohol U S S S S M S S S M S S Aluminum Chloride U U S S S S U S S S S M S S S S S S S S S M U U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S...

Page 53: ...Ether S S U U S S S U S U U S U U U U U U U U U S S S S M U Diethyl Ketone S U U M S U S M U U U M M U S S U U Diethylpyrocarbonate S S U S S S S S U S U S U S S S M S S S S S S Dimethylsulfoxide S S...

Page 54: ...S S Potassium Carbonate M U S S S S S S S S S S S U S S S S S S S S S S S S Potassium Chloride U S S S S S S S S S S S S S S S S S S S S U S S S Potassium Hydroxide 5 U U S S S S M S S S S S U S S S S...

Page 55: ...M S S M U S S M S S S M S S S S M S S S M U S M S Sodium Iodide M S S S S S S S S S S S S S S S S S S M S S S Sodium Nitrate S S S S S S S S S S S S S S S S S S S S U S S S S Sodium Sulfate U S S S S...

Page 56: ...50 U U U U S U U S M S U U M U U S S S S M S M U U M M Sulfuric Acid 10 M U U S S U U S S M U S S S S S S S S S S U U U S S Sulfuric Acid 50 M U U U S U U S S M U U S U U M S S S S S U U U M S Sulfur...

Page 57: ...Citric Acid 10 M S S M S S M S S S S S S S S S M S S S S S S S S S S 1Polyethyleneterephthalate CHEMICAL MATERIAL ALUMINUM ANODIC COATING for ALUMINUM BUNA N CELLULOSE ACETATE BUTYRATE POLYURETHANE RO...

Page 58: ...of RCF Value 4 5 F Failure message 7 3 Features 1 5 I Imbalance indicator 4 10 Intended Use v K Keys 3 2 Kundendienst 6 5 7 5 L Lid Opening 4 2 Location 2 4 M Mains Connection 2 5 Mains Supply 1 5 2...

Page 59: ...lect Program 4 9 Short term Centrifugation 4 11 Starting centrifuge program 4 10 Stopping the centrifugation program 4 11 Storage 2 6 Switch on Centrifuge 4 2 T Technical Data 1 3 Temperature Adaptati...