4
Operation
Entering Parameters
4-4
SL 40 F / SL 40 FR
Thermo Scientific
6. If available close the rotor with the rotor lid.
7. The display shows:
Entering Parameters
Acceleration curve
The Thermo Scientific SL
40
F
/ SL
40 FR offers you a total of 9 acceleration and 10 braking
curves with which samples and gradients can be centrifuged.
After the centrifuge is turned on, the last running profile selected is shown.
1. Press the ACC DEC key
once in order to open the acceleration profile selection menu.
The display shows the message
“
Set acceleration
”
.
Braking curve
1. Press the ACC DEC key
twice in order to open the braking curve selection menu.
The display shows the message
“
Set deceleration
”.
Preselecting Speed / RCF
1. Press the SPEED key
.
The display shows the
“
RPM
”
or the
“
RCF
”
-value depending on the display setting. Press the
Be sure to check all sealings before starting any aeorsol-tight applications.
See the information in the rotor instruction manual.
2. Select the desired curve by entering a digit between 1-9 on the numeric pad. Curve number 1 is
the slowest and curve number 10 the fastest.
3. Confirm your entry by pressing ENTER
.
2. Select the desired profile by entering a digit between 1-9 on the numeric pad. Curve number 1 is
the slowest and curve number 9 the fastest.
3. Confirm your entry by pressing ENTER
.