8.
(For calibrations that fail)
Select an option.
Option
Description
Details
View the calibration images.
Transfer EDS
Transfer the calibration file to Connect, a USB drive, or a network
drive (recommended for troubleshooting).
Close
The calibration file is deleted.
See “Troubleshoot calibration failure” on page 89.
9.
Unload the plate from the instrument.
CAUTION!
PHYSICAL INJURY HAZARD.
During instrument operation, the
plate temperature can reach 100°C. Allow it to cool to room temperature before
handling.
Note:
Use Smart Help if the instrument does not eject the plate (see “Request
technical support with Smart Help” on page 159).
10.
Return the plate to its original packaging.
Note:
Do not discard the packaging for the calibration plate. Each calibration
plate can be used up to three time if the following conditions are met:
·
The plate is stored in its packing sleeve at –25°C to –15°C.
·
The plate is used within 6 months after opening.
·
The plate is used before the plate expiration date.
View the calibration images
The instrument performs the ROI, uniformity, and background calibrations in
sequence. You can view the calibration images after the background calibration is
complete.
1.
In the home screen, tap
(Settings)
4
Maintenance and
service
4
Calibration
4
Status and reminders
.
The
Status and Reminders
screen is displayed.
2.
Tap a calibration type in the table, then tap
View
.
3.
Tap a calibration date in the table, then tap
View
.
The following details are displayed:
Chapter 7
Calibrate and verify instrument performance
Perform ROI/uniformity, background, and dye calibrations
7
QuantStudio
™
6 Pro and 7 Pro Real-Time PCR Systems User Guide
87