iSort
™
Automated Cell Sorter User Guide
117
Replace fluidics filters
Replace sheath inlet
filter
The 0.2-µm sheath inlet filter (Cat. No. A33307) is located in the sheath fluid line
between the sheath fluid container and the sheath pump. It should be replaced each
time sheath fluidics cleaning is performed (page 60).
Sheath fluid container
Sheath fluid line
0.2-µm sheath inlet filter
Sheath pump
1.
Detach the 0.2-µm sheath inlet filter from the bracket that secures it to the
instrument. To do this, grab the Luer lock fittings that attach the filter to the
sheath fluid line and pull it directly out.