iSort
™
Automated Cell Sorter User Guide
7.
When prompted, remove the tube containing the Attune
™
Wash Solution from
the sample tube holder, and replace with a tube containing deionized water.
8.
Move the sample tube holder up until it stops, then press
Continue
to initiate
the third step of the cleaning process.
9.
When prompted at the completion of the third step of the cleaning process,
click
Continue
to initiate the fourth step.
The instrument performs this step automatically and you do not have to
change solutions.
10.
At the successful completion of the sample line cleaning cycle, the instrument
goes back to the Cleaning Cycles screen.