Alarms
If the downflow, inflow or window parameters are not at operational settings, there may be at
risk of exposure to biohazardous material or contamination. In this case, the biological
safety cabinet emits a visual and audible alarm to alert the user, as shown in Figure 56
below.
Figure 56
Alarming Screen for Downflow too high and too low Alarm
A “Health status” icon in the shape of an alarm bell appears on a red background, along with
a “Snooze” button. Directly below the “Bell” icon, descriptive text explains the issue and
provides instructions on how to troubleshoot it. If there is more text than will fit on screen,
the text scrolls from right to left in “newsticker fashion”. When multiple alarms occur at the
same time, the “Health status” icon has a small blue circle overlay that shows the number of
pending alarms, and additional alarm descriptions appear in the newsticker section.
The symbol graphic in the Main Screen appears with a pertinent animation that identifies
the type of problem and its location. For example, the air barrier sliced in two by red
arrows shown in Figure 56 indicates that there is a downflow problem (see Figure 28 on
page 46 for a complete overview of appearances).
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...