| Start-up
36
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 5
Power Supply Connection
Establishing the power supply connection:
1. Before connecting the device to the power supply system, check to see if the voltage
of the outlet corresponds with the specifications on the nameplate of the device. If
the ratings given for voltage (V) and maximum current (A) are not correct, the device
must not be connected to the power supply system.
2. Connect the grounding plug of the device to a properly grounded and fused outlet.
The outlet must be fused separately using a fusible link T 16 A or using a circuit
breaker B 16.
3. Make sure that the power supply line is not subjected to tensile or compressive force.
Installation of the power supply connection:
To protect against accidental switch-off, the outlets for the connection to the power supply
must be located outside the normal hand reaching range and must be accessible only to
authorized personnel. Ideally, the outlets should be installed above the safety cabinet and
be readily accessible. Route the power cables so they do not run across hot surfaces
(such as exhaust tubing or similar tubing runs).
Connecting the equipotential bonding:
If the sample chamber is supplied with media (gas, water, etc.), the on-site equipotential
bonding must be connected to one of the premounted threaded bushings at the top of the
housing.
Initialization routine:
After the unit has been connected to the power supply system, the device control runs
through a start-up initialization routine.
High voltage!
Contact with current-carrying components may cause a lethal electric
shock.
Before connecting the device to the power supply system, check plug
and power supply cable for possible damage. Do not use damaged
components to connect the device to the power supply system!
The connection has to be carried out according to country specific
standards.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...