Troubleshooting
•
•
Problems and Solutions
Performance Verification fails
• If a PV wavelength accuracy test failed (peak location is outside the specified limits), use
OMNIC software to open the spectral (.SPA) file from the most recent PV test. The most
recent PV test files (polystyrene spectrum and image map) are located here:
Thermo/OMNICxi/Spectra/PV Data
–
If the spectrum contains excessive instrument noise (spikes), the polystyrene sample
may have been out of focus. Rerun the PV test making sure the focus is correct.
–
If a wavelength accuracy test fails again,
the installed laser set and
then rerun the PV test
If a wavelength accuracy test fails again, contact us.
• If the PV image resolution test failed (measured line spacing is outside the specified
limits), open the image map from the most recent PV test (see location above). The map
should contain vertical lines as in this example:
If there are no vertical lines in the collected map, check the following and then rerun the
PV test:
–
Make sure the silicon test sample is right side up (the wording on the sample should
be right side up when you are looking at the sample through the microscope)
Summary of Contents for DXR xi
Page 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Page 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Page 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Page 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Page 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Page 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Page 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Page 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...