2
Before Use
Aligning the Centrifuge
Thermo Scientific
Fresco 17 / 21
2-3
•
The set-up location must be well-ventilated at all times.
Aligning the Centrifuge
The horizontal alignment of the centrifuge must be checked every time after moving it to a different
location.
The supporting structure must be suitable for horizontal setup of the centrifuge.
Mains Connection
Figure 2-2.
Mains Connection
1. Turn off the power supply switch on the back (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
4. Make sure that the voltage and frequency correspond to the figures on the rating plate.
Establish the connection to the power supply with the connecting cable.
CAUTION
If the centrifuge is not level, imbalances can occur and the centrifuge can be
damaged.
Do not place anything under the feet to level the centrifuge.
Note
The centrifuge must be connected directly to a socket. Extensions and
connectors
are not
allowed.
Main Power Switch
Mains plug