C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
B-41
analog vout range
channel
This command reports the analog voltage output
channel
range, where
channel
is between 1 and 6 inclusive, according to
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range
channel range
This command sets analog voltage output
channel
to the range, where
channel
is between 1 and 6 inclusive, and
range
is set according to
. The example that follows sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
dig in
This command reports the status of the digital inputs as a 4-digit
hexadecimal string with the most significant bit (MSB) being input 16.
Send:
dig in
Receive:
dig in 0xff7f
din
channel
This command reports the action assigned to input
channel
and the
corresponding active state. The example that follows reports the input 5 to
be assigned an index number 9 corresponding to action of “analog outputs
to zero” with the active state being high.
Send:
din 5
Receive:
din 5 9 AOUTS TO ZERO high
set din
channel index state
This command assigns digital input
channel
(1-16) to activate the action
indicated by
index
(1-35), when the input transitions to the designated
state
(high or low). Use “list din var” command to obtain the list of supported
index
values and corresponding actions.
Table B-10.
Analog Voltage Output Range Values
Range
Output Range
1
0-1 V
2
0-100 mV
3
0-10 V
4
0-5 V
0 [cannot be set to this, but may report]
Undefined