48
Installation, Operation and Service Manual
Thermo Scientific XBF40-MD Series Blast Freezers
Appendix
Regulation
CCt
Continuous
cycle duration
0.00
Hours
Regulation
CCS
Set point for
continuous
cycle
0
-100
150
°C
Regulation
Con
Compressor ON
time with faulty
probe
255
0
255
Minutes
Regulation
CoF
Compressor OFF
time with faulty
probe
0
0
255
Minutes
Regulation
CF
Temperature
measurement unit
°C
Regulation
rES
Resolution
in
Regulation
Lod
Probe displayed
P1
Regulation
rEd
X-REP display
P1
Regulation
dLy
Display temperature
delay
0.00
Minutes
Regulation
dtr
P1-P2 percentage
for display
99
1
99
Defrost
EdF
Defrost mode
rtC
Defrost
tdF
Defrost type
EL
Defrost
dFP
Probe selection for
first defrost
P2
Defrost
dSP
Probe selection for
second defrost
P2
Defrost
dtE
Defrost termination
temperature first
defrost
10
-55
50
°C
Defrost
dtS
Defrost termination
temperature second
defrost
10
-55
50
°C
Defrost
idF
Interval between
defrost cycles
8
0
120
Hours
Defrost
MdF
(Maximum) length
for first defrost
30
0
255
Minutes
Defrost
MdS
(Maximum) length
for second defrost
30
0
255
Minutes
Table 4: Dixell Control Parameter and Defaults