8
•
Troubleshooting
Page 212
Integral Fraction Collector FT (VF-F20-A) Operating Manual
8.3
Operating Issues
This section gives an overview of possible operating issues and remedial
actions.
8.3.1
Resolving Front Leakage
When
The leak sensor is wet. The leak sensor reports leakage.
Parts required
Replacement part as required
Additional items required
Cloth or tissue
Preparations
When resolving leakage, observe the safety guidelines and general rules
for maintenance and service as presented in
(
page 142)
.
Follow these steps
1. Locate the source of the leak.
As leakage usually occurs at a connection, visually inspect all
components and connections in the flow path.
2. Tighten or replace the connection or component as required.
3. With a cloth or tissue, thoroughly absorb all liquid that has collected
in the leak tray and under the leak sensor. Be careful not to bend the
sensor.
4. Allow the sensor to adjust to the ambient temperature for a few
minutes.
5. If leakage is no longer reported, you can resume operation.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...