7.4.2
Flushing the Pump
When
•
To avoid contamination of the pump, see
section 7.4.2.1
•
With persistent contamination, see
7.4.2.1
Avoiding Contamination
When
To avoid contamination of the degassing channels
Items required
Fresh solvent
Observe the following guidelines
The following guidelines refer to conventional use of the Vanquish
system as recommended in this manual, such as using fresh high-quality
solvents (minimum UHPLC-grade or LC/MS-grade) that are replaced
frequently:
•
Regularly flushing the degassing channels with fresh solvent is
usually sufficient.
•
If you use water, acetonitrile or methanol for your chromatography,
flushing the channels once a week is an appropriate interval.
•
Mobile phases containing salts, buffering agents and/or other
additives are inherently prone to organic contamination and/or
biodegradation. They may require a much shorter flushing interval.
•
Adapt the interval to the solvents in use.
TIP
With persistent contamination, for example, if ghost peaks appear
reproducibly in your blank chromatogram without injecting a sample,
see the next section in this manual.
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...