7
•
Maintenance and Service
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 119
7.3
General Rules for Maintenance and Service
For successful maintenance and service procedures, follow these rules
and recommendations:
•
Before starting maintenance or service procedures, shut down the
pump when instructed to do so.
•
Use only the replacement parts specifically authorized and qualified
for the pump by Thermo Fisher Scientific. For ordering information,
see
section 10.4 Consumables and Replacement Parts, page 223
•
Follow all instructions step by step and use the tools recommended
for the procedure.
•
Before opening the flow path to replace capillaries in the system,
turn off the pump flow and wait until the system pressure is down to
zero.
•
Dirty components can contaminate the chromatography system.
Contamination leads to poor performance of the modules and entire
system or can even cause damage to the modules and system.
Therefore:
♦
Always wear appropriate gloves.
♦
Place the components only on a clean, lint-free surface.
♦
Keep your tools clean.
♦
Use only lint-free cloth for cleaning.
•
After you have performed maintenance or service on the flow
connections in the pump, test the pump for leakage before you
resume operation.
•
If you need to return the pump for depot repair, follow the
instructions in
section 7.16 Transporting or Shipping the Pump,
page 186
Summary of Contents for Vanquish VH-P10-A-02
Page 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Page 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...