Test
Procedures
OQ/PQ Operating Instructions
M1393R0
Page 31 of 60
4.7
Linearity of the UV Detector
The detector linearity is measured by injecting the different caffeine standards of the
OQ/PQ kit (P/N 16395) as presented in TABLE 3-5. The resulting peak height is used for
the calculation. An example of an injection (0.2
L) of a caffeine standard of 20 µg/mL
onto a Nano UltiMate system is shown in FIGURE 4-7.
FIGURE 4-7
UV Trace of the Injection of Caffeine onto a Nano UltiMate System
An example of a calibration curve is shown in FIGURE 4-8.
FIGURE 4-8
Example of a Calibration Curve -
Nano LC system, 99.93 % Correlation Coefficient