15
Calibrating the MS
Determining the Initial API Source Settings
Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
57
Figure 25.
Syringe pump settings box
c.
Click
Apply
.
5. Verify that the inlet plumbing connections do not leak.
6. Open the syringe pump settings box again. Press and hold
Prime
to prime the syringe at
100 μL/min.
7. Verify that the system readback is normal,
.
Determining the Initial API Source Settings
These initial setting are a starting point for calibrating and optimizing system performance.
1. In the Tune window, click the
Ion Source
tab.
2. In the Current LC Flow (μL/min) box, type the flow rate from
, and then click
Get Defaults
.
3. Click
Apply
.
The Tune application makes a change in the History pane.
Table 11.
Recommended LC flow rates
Solution Type
Injection Type
LC flow rate (μL/min)
Calibration
Syringe
5
Sample Syringe
3-10
Autosampler
200