Restarting the MS System After a Shutdown
When you restart the MS system after a complete shutdown, the sequence of tasks is
important.
• Resetting the Mass Spectrometer
• Resetting Calibration Parameters
• Starting the LC, Gases, and Autosampler
Restarting the Data System
If possible, use the Windows 7 or Windows 10 restart procedure the data system so that
Windows can properly close applications and save changes to any open application.
Restarting the Data System Using Windows
1. On the Windows taskbar, choose
Start
, and then click the arrow next to Shut Down.
2. Choose
Restart
, and then click
OK
.
Restarting the Data System Using the Power Button
1. Press the power button on the data system computer.
2. Wait at least 20 seconds.
3. Press the power button again.
Note
After you reset the data system, the communications link between the data system
and the MS is automatically reestablished. When this occurs, the Communication LED
turns yellow and then green. If the system is unable to reestablish the communications
link, follow
“Resetting the Mass Spectrometer.”