Advanced Troubleshooting
Replacing the Calibration Gas Components
376
TSQ 9000 Hardware Manual
Thermo Scientific
Calibrant Reservoir Kit
To replace the calibrant reservoir kit, which consists of the calibrant reservoir and its
cover
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Unscrew the hydrogen safety screw.
3. Open the front door of the TSQ 9000 instrument.
4. Twist the calibrant reservoir cover counter-clockwise and remove it from the calibration
gas module.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components. If you just need to refill the calibrant reservoir and not replace it, see
Refilling the Calibrant Reservoir
Note
You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 400.
Removing the Calibrant Reservoir Cover
Calibration
Gas Module
Calibrant
Reservoir Cover