Performing Routine Maintenance
Maintaining the Calibration Gas Module
70
TSQ 9000 Hardware Manual
Thermo Scientific
Refilling the Calibrant Reservoir
To refill the calibrant reservoir:
1. Open the front door of the TSQ 9000 instrument.
2. Twist the calibrant reservoir cover counter-clockwise and remove it from the calibration
gas controller.
3. Fill a syringe with 200 μL of the FC 43 calibration compound, which contains
perfluorotributylamine (PFTBA), and insert it into the calibrant reservoir.
Note
It is not necessary to vent the instrument to perform this operation
Figure 62.
Removing the Calibrant Reservoir Cover
Figure 63.
Removing the Calibrant Reservoir from the Cover
Calibration
Gas Controller
Calibrant
Reservoir Cover
Calibrant
Reservoir
Cover
Calibrant
Reservoir