Advanced Troubleshooting
Replacing the Covers of the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
295
Right Side Panel
To replace the right side panel of the TSQ 9000 instrument
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the column from the transfer line completely.
3. Slide the top cover panel toward the back of the instrument and lift it off.
4. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
5. Replace the right side panel and tighten the screws.
6. Replace the top cover panel and tighten the screw holding it in place.
7. Reinsert the column to the proper length and tighten the transfer line nut. See
for more information.
8. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge. However, you must vent the
instrument and remove the column.
Figure 316.
Removing the Right Side Panel Screws
Right Side
Panel Screws