Medical Devices -Touch Controller
13
| Transport and Handling
Indication for Optimal Use
The following instructions are provided to the user to
follow in order to get the best equipment performances:
-
The power supply must comply with
the information provided on the
technical data plate (+/- 10%).
-
The units have been designed and
built to work in environment with
temperatures falling within the climatic
class temperatures indicated on
the technical data plate (see
) and at a relative
humidity of 60%.
-
Do not block the motor compartment
air vents.
-
Load stored material gradually at
ambient temperature to grant proper
refrigeration.
-
Store material on drawers.
Do not place products directly on the
bottom or against the wall, doors or
fixed guards of the unit (see
Material Inside the Compartment
)
-
Ensure doors are closed properly.
-
Limit door opening frequency and
duration. Each time the door is
opened, the internal temperature will
alter and there will be possible ice
formation on the evaporator
-
Keep the defrost water drain outlet
clear.
-
Follow a regular maintenance
schedule (see
).
Every model of freezer is designed to store
material already frozen, especially model
SUPERARTIC. Material load turnover
cannot exceed 5% of the total stored load.