3
Setup Menu Item
Setup Menu Item Modules
148
TriPlus RSH User Guide
Thermo Scientific
Table 70.
Tray Type Parameters (Sheet 1 of 3)
Item
Description
1st Sample Offset X
Offset distance from
Tray Zero Point
to first position in
X-axis direction. Unit: mm.
1st Sample Offset Y
Offset distance from
Tray Zero Point
to first position in
Y-axis direction. Unit: mm.
1st Sample Offset Z
Offset distance from
Tray Zero Point
to first position in
Z-axis direction. Unit: mm.
(This parameter is currently not used with released Tray Types. Value = 0)
Alt Sample Offset X
Offset from
Tray Zero Point
to first sample position of
staggered row/column in direction X-axis. Unit:
μ
m.
Alt Sample Offset Y
Offset from
Tray Zero Point
to first sample position of
staggered row/column in direction Y-axis. Unit:
μ
m.
Alt Sample Offset Z
Offset from
Tray Zero
Point
to first sample position of
staggered row/column in direction Z-axis. Unit:
μ
m.
(This parameter is currently not used with released Tray Types. Value = 0)
Cavity Depth
Defines the total depth of the cavity. Unit: mm.
Cavity Diameter
Defines the diameter of the cavity. Unit: mm.
Columns
Number of columns.
Indexing Orientation
Defines the order of sampling either in
Column Direction
or in
Row direction
.
Examples:
Tray R60: Column Direction. Tray VT54: Row
Direction.
Length X
Total length of the Tray in X-axis direction, outer frame of
Tray is considered. Unit: mm.
Length Y
Total length of the Tray in Y-axis direction, outer frame of
Tray is considered. Unit: mm.
Length Z
Total length of the Tray in Z-axis direction, outer frame of
Tray is considered. Unit: mm.
Penetration Vial Index
Specifies the position in the Tray where the cut-out for
needle penetration adjustment is located.
Rows
Number of rows.