Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
141
Naming Convention and Relationship between Software Classes‚ Tray Holder,
Trays, Tray Types, and Cap Type
The RSH Firmware filters the selection of the subsequent classes based on the selection of the
initial item. You should be aware of several important points.
First, the
Needle Guide Type
must be defined. If
Not Specified
is selected under
Needle
Guide Type
, the execution of the run will be prohibited.
In addition, the
Cap Type
and
Needle Guide Type
must be compatible in order to transport
a vial. If a mismatch is observed, such as if a 2 mL vial with a magnetic cap is in use but the
Needle Guide Type is selected for a 20 mL vial, then the software will prohibit the execution
of the run.
illustrates the interdependency between the various software classes and subclasses.
The different colors depict the main classes and the parameters assigned to each of them.
See
.
Figure 93.
Relationship between Software Classes: Tray Holder, Tray, Tray Type, Cap Type, and
Needle Guide Type.
Color Coding
The software classes are labeled in
red
. All related software classes where a
Vial
is used are
displayed using a
green
shade and the classes related to
Well Plates
are displayed using a
blue
shade.