3
Setup Menu Item
Setup Menu Item Modules
134
TriPlus RSH User Guide
Thermo Scientific
Tool Adapters
Selecting the list item
Tool Adapters
opens up the following parameters. See
Figure 89.
Menu Item Tool Adapter and List Item MHE Station
Selecting the list item
Tool Adapters
opens the dialog window for
MHE Station
.
Press
Enter
to edit a parameter. See
Tool Adapter Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Table 64.
MHE Station Adapter Parameters
Item
Description
Z-Tolerance
A tolerance window to give a plus/minus range (expressed in
mm) where the Head must expect an object.
If the value of this item is
> 0
, this indicates a relative
detection mode, and the sensor from the Head needle guide
is active.
If the value of this item is set to
0
the Needle Guide Motor
detection current is turned off and the syringe slider moves
to an absolute value, the Z-axis position as specified by the
object.
Table 65.
Tool Adapter Options Parameters (Sheet 1 of 2)
Item
Description
New MHE Station
Creates a new MHE Station.
Copy MHE Station
Copies the present MHE Station.