TriPlus RSH Basic Informations
Operation of the TriPlus System in Combination with Other Chromatography Data Systems
92
TriPlus RSH User Guide
Thermo Scientific
Figure 71.
Menu Item Shutdown
The main advantage of this automated shut down is that the Tool is parked. If the system is
shut down by simply turning off the power supply, the slider will move very slowly
downwards.
Parts of the Tool, particularly the syringe needle, can touch the bench surface, or any other
object. This can damage the syringe needle.
The following steps are automatically performed during a controlled Shutdown:
• The Head moves to an empty position of the ATC Station, engages and remains in this
position.
• In case a TR Station is part of the TriPlus RSH configuration, the mead moves to the
single slot.
• The software shuts down, closes the application and disables the axes.
After the TriPlus RSH has completed the shutdown procedure, it is necessary to turn off the
power at the power supply (switch at power).
To restart the TriPlus RSH after the shutdown process, toggle the switch of the power supply,
turn it off – wait about 30 seconds – then turn on the power of the system.
Operation of the TriPlus System in Combination with Other
Chromatography Data Systems
For operation of the TriPlus RSH with a Thermo Scientific Chromatography Data System
(CDS), for example: Chromeleon, Xcalibur, ChromQuest, or Chrom-Card, a software driver
is used. The specific requirements for operating the integrated System within the CDS are
outlined by the supplier of the CDS software.