Chapter 14
Replace the Interferential Filter
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
297
OPERATING SEQUENCE
Replace the Interferential Filter
Materials needed
•
Detector fixing tool (P/N 205 021 50)
•
Interferential filter (see Table 14-2 for available filters)
•
Paper towels
Table 14-2.
Available interferential filters for the FPD
Target compound
P/N
Phosphorus
281 071 00
Sulphur
281 070 00
Organotin
281 070 01
detector
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS