Chapter 12
Maintaining an Electron Capture Detector
ECD Troubleshooting
236
Maintenance and Troubleshooting Manual
NOTE
When packed columns are used, the make-up gas (as carrier gas type) is used only if
strictly required.
WARNING! It is the customer’s responsibility to ensure compliance with all local safety
regulations concerning gas supplies.
Hydrogen is a potentially dangerous gas. Mixed with air it may give rise to an
explosive mixture. The use of hydrogen requires the operator’s extreme caution due
to the risks involved. For further details concerning hydrogen, please refer to
Contamination
Detector contamination could be indicated by the following effects in the
chromatogram:
•
reduced signal to noise ratio
•
high-noise baseline
•
baseline drift with changing pulse voltage
•
negative dips after peaks