Troubleshooting
Investigating Power Supply Issues
582
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Investigating Power Supply Issues
Verify the correct power supply to the instrument.
TRACE 1300/TRACE 1310 will not power-on
Investigating Communication Issues
Verify the instruments is communicating with the computer.
Software is not communicating with the TRACE 1300/TRACE 1310
TRACE 1300/TRACE 1310 does not start or is not ready
Cannot download methods to the TRACE 1300/TRACE 1310
Possible Remedies
Make sure the TRACE 1300/TRACE 1310 power cable is properly connected to the instrument and to the correct 220/120 V main power
line outlet.
Verify that the electrical outlet is functioning properly.
Power supply system is faulty. Contact your local Thermo Fisher Scientific customer support organization.
Possible Remedies
Make sure the LAN cable is properly connected to the GC.
Confirm the TCP/IP configuration on the computer matches the GC.
Restart the GC.
Possible Remedies
Verify the GC methods and configuration.
Make sure the electrical connections have been properly carried out.
Confirm the GC handshaking parameters are set properly.
Possible Remedies
Verify that your instruments are properly configured.