11
Upgrade Equipment
Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
561
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
10. Tune the ISQ Series / TSQ 8000 Series and set its working conditions. Set the GC
working conditions accordingly.
For details please refer to the
TRACE 1300 and TRACE 1310 User Guide
, and to the
ISQ
or TSQ 8000 Series User and Hardware manuals
.
Coupling with the DSQ II Mass Spectrometer
To update the GC for the coupling with the DSQ II mass spectrometer
1. Make sure that the preliminary operations have been carried out. See
2. Remove the left side panel.
a.
Open the front door of the GC.
b. Use a T20 Torxhead screwdriver to screw the left side panel screw from the interior
front panel. See
. Save the screw because it will be reused later.
Figure 615.
Left Side Panel Fixing Screw
c.
Slide the panel towards the back of the instrument up to the stop.