1
Installation
Installing the Injector and Detector Modules
6
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
d. Close the module flap cover.
3. Continue the installation following the instructions reporting in the section
Gas Supply Plumbing Connections”
Installing a Detector Module
To install a detector module
1. Remove the dummy module from the position where the detector module will be
installed.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the two captive fixing screws.
c.
Keeping the dummy module flap cover open, lift up the module from its seat in the
injector/detector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver.
Figure 3.
Detector Gas Block Plug
CAUTION
If you are installing a
NPD detector module
, the installation of the NPD
Thermionic Source Power Module is required. For the installation details see the section
“Adding a NPD Detector Module”
.
CAUTION
If you are installing a
Generic Detector Interface
see the section
.
Gas Connections Without Gas Block Plug
Gas Block Plug on Detector Seat
O-rings