Performing Injectors Routine Maintenance
Maintaining a Programmable Temperature Vaporizing Injector with Backflush (PTVBKF)
128
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 109.
PTVBKF Injector Components
The PTVBKF injector periodic maintenance includes:
• Replacing the PTVBKF septum
Replace the septum at least after every 200 injections, or every time a problem related to
septum damage, or wear occurs.
See
.
Septum Cap
Septum
Liner Cap
Liner Seal
Liner
Split Nut
Silver Seal
Ferrule
Terminal Fitting for Capillary Column
Backflush Line
Backflush Line Fitting
Ferrule
Fitting Cap, Backflush Line