8
•
Troubleshooting
Page 210
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
8.5
Pressure Issues or Shifts in Retention Time
When you observe pressure issues or shifts in retention time, the
following sections help you to identify and eliminate the source for the
problem:
•
The pressure is
not
down to zero when the purge valve is open
Pressure Not Zero with Open Purge Valve (
.
•
Pressure pulsation
See
Checking the Compression Values (
Static Mixer for Permeability (
•
Shift in retention time
See
Checking the Compression Values (
.
•
Inconstant pressure or high backpressure at the column and pump
See
Testing the Static Mixer for Permeability (
8.5.1
Pressure Not Zero with Open Purge Valve
When
The pressure is
not
down to zero when the purge valve is open.
Follow these steps
1. For the pressure sensor for which the pressure is not down to zero,
check the pressure sensor cable and connector for proper contact.
a) Unplug the pressure sensor cable.
b) Reconnect the cable. To ensure proper contact push the cable
firmly into the connector.
2. Perform a self-test.
3. If the self-test produces a message, follow the remedial actions
recommended for the message.
8.5.2
Checking the Compression Values
The compression values of the pump heads can provide valuable
information for troubleshooting.
When
You observe pressure pulsation or shifts in retention time.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...