7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 139
7.6.2.1
Removing the Pump Head
Preparations
Prepare the pump for removing the pump head. See
Follow these steps
Figure 37: Pump head
No.
Description
1
Pump head inlet
2
Pump head screw
3
Pump head outlet
4
Seal wash tubing
1. Open the purge valve by turning the valve knob counterclockwise.
2.
As required by the pump type:
Pump Type
Action
Binary pump
Select the pump head for which you want to
perform the procedure.
Dual pump
Select the pump for which you want to perform
the procedure.
Isocratic pump, quaternary pump
Continue with the next step.
3. Undock the pistons either from the keypad or from the Chromeleon
software. Wait until undocking is complete before proceeding with
the next step.
4. Remove the capillary that connects the pump head to the purge/
sensor unit. Disconnect the capillary on both the pump head outlet
and the purge/sensor unit.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...