7
•
Maintenance and Service
Page 138
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
7.6.2
Replacing the Pump Head
When
•
Damage of pump head
•
Problems with pump performance
Parts required
•
Pump head
•
Spacing tool (pump head tool)
Tools and additional items required
•
Hexagon wrench, size 6
•
Depending on the capillary that connects the pump head to the
purge/sensor unit:
¨
Viper capillary: Installation tool for Viper capillaries with torque
toothing
¨
Conventional capillary (non-Viper): Open-end wrench, size 1/4"
•
Solvent suitable for purging to flush out harmful substances
•
Seal wash liquid
•
Solvent suitable for your application
Preparations
1. To flush out harmful substances, purge the pump with an
appropriate solvent.
2. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
1. Remove the pump head. See
.
2. Install the pump head. See
.